Operation
▬▬▬▬▬▬▬▬▬▬▬▬▬▬▬▬▬▬▬▬▬▬▬▬▬▬
SafeCheck 200 Operator’s Manual
15
The other option in the DISPLAY OPTIONS menu is LEL. This allows
you to select the way that the gas reading from the LEL (Combustible)
sensor will be displayed. With LEL highlighted, pressing I/O will change
between % of LEL or % of CH
4
by volume. As % of LEL it will display as
a percentage of the Lower Explosive Limit from 0 to 100%. As % of CH
4
it will display as a percentage of the volume of CH
4
(methane) that is
present in the environment from 0.0 to 5.0%. Note: 5,0% by volume of
CH
4
is also 100% LEL for CH
4
. Press the arrow key to return to the
setup menu.
• Password:
This menu item allows you to set or change your unit's password. The
password is necessary to enter the setup mode and change any of the unit's
setup parameters.
Use the arrow key to move through the setup menu until PASSWORD
is highlighted, then press I/O. Enter the new
password using the arrow key to change the
digits and the I/O key to set them. After the last
digit is entered, you will be asked to re-enter the
same password as a check. If they match, the display will show
PASSWORD ACCEPTED and then return to the setup menu. If the re-
entered password does not match the initial one, you will be asked to begin
again. Pressing the arrow key at any time will return to the main setup
menu.
3.4.4 Calibration:
This should be done for each installed sensor prior to entering your work
area. An uncalibrated unit can be as dangerous as no unit at all. Failure to
calibrate can result in inaccurate sensor readings.
From the main menu, press the I/O key while CAL. is highlighted The
following items, will be found in the CAL.
menu: INDIVIDUAL CAL, ALL CAL, and
SPAN ADJUST. Press the arrow key to cycle
through them. Press the I/O key to select the highlighted item.
Enter New
Password: ---
Individual Cal
All Cal
Operation
▬▬▬▬▬▬▬▬▬▬▬▬▬▬▬▬▬▬▬▬▬▬▬▬▬▬
SafeCheck 200 Operator’s Manual
16
• Individual Cal: The unit first waits to do a CALIBRATION ZERO. This
is automatically done before any calibration to
adjust for drift in the sensor's sensitivity, which
occurs over time and after use.
Note
: Fresh uncontaminated air is necessary for the proper outcome of
this procedure. (Zero the unit away from heavy traffic, excessive
cigarette smoke, and industrial fumes.) If such an atmosphere is
difficult to reach, zero air cylinders are available.
Press the I/O key when you are sure that you are in a clean air environ-
ment. The unit goes through and automatically zeros all the installed sen-
sors.
The installed sensor list now appears in the top line of the display. The
second line gives you the date of the last cali-
bration for the highlighted sensor. Use the
arrow key to scroll through the sensors. Press I/O
to access the highlighted gas calibration menu.
At this point the unit takes over and runs the calibration. On the display,
the unit asks you to apply the concentration of
calibration gas that was specified in Span
Adjust on page 19. The second line displays
the message "Waiting......" along with the current concentration.
Place the calibration cup over the unit's sensors on the SafeCheck 200.
Screw the regulator onto the calibration gas cylinder.
Connect tubing between the regulator on the gas cylinder and the calibra-
tion cup on the SafeCheck 200. Open the regulator as far as it will turn and
gas should begin flowing over the sensors. Once the level of gas begins to
rise, the second line of the display will change to "Detected" and the
number next to it will begin to rise. Once the rise in the gas level starts to
slow down and reach a stable level, the unit displays the message
"Stabilizing". When the unit determines that the gas level has stabilized
and no longer rising, the unit will do a calibration. It will display a mes-
sage "Calibrating" and if successful "Calibrated". If calibration is not suc-
cessful, the unit will indicate "Cal Failure". Check your calibration setup
and calibration gas and then attempt another calibration. If the unit still
cannot calibrate, either replace the sensor or have the unit serviced.
Calibration Zero
Ready?
O2 LEL
Apply O2 @ 20.9
Waiting …. 18.0